Standard

ISO 5347-13:1993

Published

Corrigendums and amendments are bought separately.

Language
Services

Abstract

Lays down detailed specifications for the instrumentation and procedure to be used for base strain sensitivity testing. Applies to rectilinear accelerometers, mainly of the piezoelectric type. Is applicable for the following parameters: reference values - pick-up base radius of curvature: 25 m, base strain: 2,5 × 10^-4; check values - radius of curvature 62,5 m, strain of base 1 × 10^-4; radius of curvature 12,5 m, strain of base 5 × 10^-4.

Document information

  • Standard from ISO
  • Published:
  • Edition: 1
  • Version: 1
  • Document type: IS
  • ICS 17.160
  • ISO TC ISO/TC 108