Standard

ISO 5347-17:1993

Tilbaketrukket

Rettelser og tillegg kjøpes separat.

Språk
Tjenester

Omfang

Lays down detailed specifications for the instrumentation and procedure to be used for fixed temperature sensitivity testing. Applies to rectilinear pick-ups. Is applicable for the following parameters: frequency range: 20 Hz to 1250 Hz; dynamic range: 0,1 µm to 10 mm (frequency-dependent), 1 mm/s to 1 m/s (frequency-dependent), 10 m/s^2 to 1000 m/s^2 (frequency-dependent); temperature range: - 45 °C to + 800 °C. The uncertainty applicable is ± 10 % of reading.